![]() |
DETECTABUSE™ GRAVITY SERIES GV-65 CATION EXCHANGE METHOD FOR THE ANALYSIS OF TRICYCLICS IN URINE BY GC-MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
|
Revised: March, 2003 Supercedes May, 2001 - Revision addresses stability of TEA in sample elution section |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Add 2.0 mL of urine to a 16 x 100 mm
disposable borosilicate glass tube with an inert screw cap top.
Note: When adding an internal standard dissolved in an organic
solvent to a urine or blood sample, the solvent volume must not
exceed 3% of the buffered sample volume. Higher solvent
concentrations may produce extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity.
1. Pour samples onto preconditioned column. Allow to flow by
gravity. Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there
is either air trapped within the column bed or frits, or there is aqueous
phase remaining on the column because of a weak vacuum during the column
drying step. Tapping the column mounting plate onto the vacuum box should
initiate flow. If this does not do the job, use a rubber bulb to
gently push a few drops of elution solvent and trapped air into the collection
tube. Allow the remainder of solvent to flow by gravity.
1. Sample elution is done outside of the vacuum box. DERIVATIZATION - MBTFA
1. To each dried extract add 50 µL Acetonitrile, vortex mix, then add
50 µL MBTFA. Note: Polar solvents commonly used for derivatization
such as Acetonitrile or Ethyl Acetate will pick up moisture over time.
Because moisture will inhibit or prevent derivatization it is important
to keep a supply of solvents used for this purpose stored as protected
from moisture as possible. Storing these solvents in a desiccator or flushing
with Nitrogen or Argon after use is recommended.
SUPPLEMENT - When using an automated robotic system all liquids may be allowed to flow unassisted through the column or may be pulled through the column with vacuum or pushed through with positive pressure.
|
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 265°C |
| Detector Temp.: 300°C |
| Initial: 140°C, program
@ 200C/min. to 245°C Final: 245°C-285°C @ 30C/min |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 2000C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 4.0 min. |
| Start Acq.:4.0 min. |
| Stop Run:8.0 min. |
| MSD PROGRAM | ||
| Drug | Ions Monitored | Retention Time |
| Amitriptyline | 189, 202, 215 | 5.56 |
| Trimipramine | 234,248, 249 | 5.67 |
| Imipramine | 234, 235,280 | 5.69 |
| Nortriptyline | 202, 204, 232 | 6.45 |
| Clomipramine | 268, 269, 270 | 6.53 |
| Protriptyline | 191, 192, 359 | 6.61 |