DETECTABUSE™ GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF PSILOCIN IN URINE BY GC-MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: October, 2003 |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Add 3.0 mL of urine to a 16
x 100 mm disposable borosilicate glass tube with an inert screw cap top. Note: When adding an internal standard dissolved in an organic solvent to a urine or blood sample, the solvent volume must not exceed 3% of the buffered sample volume. Higher solvent concentrations may produce extraction losses. HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions) COLUMN CONDITIONING 1. Wash column with 1 mL of Methanol. Allow to flow by gravity. SAMPLE EXTRACTION - (Please see Notes at end of this section before proceeding) 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there is probably air trapped within the column bed or frits. Tapping the column mounting plate onto the vacuum box should initiate flow. Any columns that have not emptied within 5 or 6 min. may be induced with a low vacuum from a small vacuum pump. SAMPLE ELUTION 1. Sample elution is done outside of the vacuum box. DERIVATIZATION 1. To each dried extract add 75 µL Acetonitrile, vortex mix, then add
25 µL MSTFA.
Assisted flow parameters may be set as follows:
|
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 265°C |
| Detector Temp.: 285°C |
| Initial: 130°C, program at
20°C/min. to 240°C |
| Final: program at 30°C/min.
to 280°C Hold for 1 Minute |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell:20 |
| Solvent Delay: 4.5 min. |
| Start Acq.:4.5min. |
| Stop Run: 8.0 min. |
| References:
F.Hasler, D.F. Bourquin, R.M. Brenneisen, T. Bar, and F.X. Vollenweider, Urinary excretion Profiles of Psilocin After Oral Administration of Psilocybin. J. Anal. Toxicol., 21 (1997) |
| MSTFA | ||
| Drug | Ions Monitored | Retention Time |
| Psilocin-D10 | 66, 358 | 5.33 |
| Psilocin | 58, 290, 348 | 5.35 |