![]() |
DETECTABUSE™ GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF KETAMINE IN URINE BY GC/MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on it's performance and welcome your suggestions for improvements or enhancements. |
Revised:January, 2001 |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Add 1.0 mL of urine to a 16 x 100 mm disposable borosilicate
glass tube with an inert screw cap top.
Note: When adding an internal standard dissolved in an organic
solvent to a urine or blood sample, the solvent volume must not
exceed 3% of the buffered sample volume. Higher solvent
concentrations may produce extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity.
SAMPLE EXTRACTION - (Please see Notes at end of this section before proceeding)
1. Pour samples onto preconditioned column. Allow to flow by
gravity. Samples will flow through the column at a rate of 1-2 mL/min.
Note: If liquids do not elute freely by gravity flow, there is probably air trapped within the column bed or frits. Tapping the column mounting plate onto the vacuum box should initiate flow. Any columns that have not emptied within 5 or 6 min. may be induced with a low vacuum from a small vacuum pump.
SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box.
DERIVATIZATION (Only required if being assayed with Amphetamines)
Using HFBA:
WITHOUT DERIVATIZATION
1. Add 100 µL of Ethyl Acetate to each tube.
SUPPLEMENT - When using an automated robotic system
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 1750C |
| Detector Temp.: 2500C |
| Initial: 750C, Hold for 0.5 min., program at 200C/min. to 2050C |
| Final: 2750C at 400C/min. |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 2000C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Solvent Delay: 4.0 min. |
| Dwell: 30 |
| Start Acq.:3.0 min. |
| Stop Run: 9.0min. |
| MSD SIM PROGRAM | ||
| Underivatized | ||
| Drug | Ions Monitored | Retention Time |
| Norketamine | 166, 168, 195 | 6.91 |
| Ketamine | 180, 182, 209 | 7.12 |
| Ketamine-d4 | 184, 186, 213 | 7.10 |
| HFBA derivatization | ||
| Drug | Ions Monitored | Retention Time |
| Norketamine | 340, 356, 384 | 6.80 |
| Ketamine | 210, 362, 370, | 7.61 |
| Ketamine-d4 | 366, 374 | 7.60 |
![]() GO TO: GV-65 Ordering Information | ||