BCDlogo

DETECTABUSE™ GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF
BARBITURATES IN URINE BY GC-MS


Although this method has performed well in our laboratory, it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements.

Revised: March, 2003
Supercedes May, 2001 -
Revision addresses stability of TEA in sample elution section

SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)

1. Pipet 1.0 mL of urine to a 16 x 100 mm disposable borosilicate glass tube.
2. Add 150 ng of Phenobarbital-d5 per mL of sample as internal standard..
3. Add 3.0 mL of 0.25M Phosphate Buffer, pH 9.1. Vortex mix.
4. If adjusted sample is turbid or precipitated centrifuge for 3 minutes at 3000 RPM.

Notes: When adding an internal standard dissolved in an organic solvent to a urine or blood sample, the solvent volume must not exceed 3% of the buffered sample volume. Higher solvent concentrations may produce extraction losses.

HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)

COLUMN CONDITIONING

1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity.
2. Proceed to Sample Extraction within 30 min. of column conditioning.

SAMPLE EXTRACTION - (Please see Notes at end of this section before proceeding)

1. Pour samples onto preconditioned column. Allow to flow by gravity. Samples will flow through the column at a rate of 1-2 mL/min.
2. Add 3.0 mL of Deionized Water to each column. Allow the columns to flow by gravity.
3. Add 1.0 mL Water:Methanol (60:40) to each column. Allow to flow by gravity.
4. Dry the columns by applying vacuum adjusted to at least 7" Hg for 5 minutes. (Test by momentarily placing the heel of hand over the column top. A strong pull should be felt through the column).

Note: If liquids do not elute freely by gravity flow there is probably air trapped within the column bed or frits. Tapping the column mounting plate onto the vacuum box should initiate flow. Any columns that have not emptied within 5 or 6 min. may be induced with a low vacuum from a small vacuum pump.

SAMPLE ELUTION

1. Sample elution is done outside of the vacuum box.
2. Place the column mounting plate on the elution rack loaded with an appropriate number of 12 x 75 mm or 15 x 85 mm borosilicate glass test tubes. Make sure that the hole pattern on the plate matches the hole pattern on the rack.
3. Add 1.0 mL of basic n-Butylchloride containing 4% (v/v) Triethylamine (TEA)* to each column and allow solvent to flow through the columns by gravity into the test tubes.
4. Dry under N2 or argon at less than 60°C.

Note: If liquids do not elute freely by gravity flow, there is either air trapped within the column bed or frits, or there is aqueous phase remaining on the column because of a weak vacuum during the column drying step. Tapping the column mounting plate onto the vacuum box should initiate flow. If this does not do the job, use a rubber bulb to gently push a few drops of elution solvent and trapped air into the collection tube. Allow the remainder of solvent to flow by gravity.

* Elution solvent with 4% TEA, (4 mL TEA is added to 96 mL of n-Butyl Chloride) is stable for approximately one week stored in a glass bottle with a Teflon or polypropylene lined cap. Close bottle tightly when not in use. A white residue begins to appear in the dried down eluate when the TEA begins to deteriorate. Artifacts from this process may interfere with "fast" GC/MS methods.

ON-COLUMN DERIVATIZATION

1. Dissolve residue in 100 µL of 0.2M Trimethylphenyl Ammonium Hydroxide (TMPAH) in Methanol.
2. Cap the tubes or transfer into 100 µL reaction vials and seal.
3. Inject 2 µL into GC-MS for on-column derivatization.

SUPPLEMENT - When using an automated robotic system all liquids may be allowed to flow unassisted through the column or may be pulled through the column with vacuum or pushed through with positive pressure.
Assisted flow parameters may be set as follows:
Column Conditioning - Pass through column in approximately 20 seconds (± 20%).
Sample, Sample Washes, and Elution Solvent - Pass through column in approximately 60 seconds (± 20%).
Column Drying Steps - Use 12-15 PSI of positive pressure for 40 seconds or vacuum set at 15" Hg for 30 seconds (These drying parameters are for individual columns).


GC/MS ANALYSIS
GC/MS: Hewlett-Packard equipped with Mass Selective Detector
GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film
Acquisition Mode: SIM

Temperature Program:
Injector Temp.: 280°C
Detector Temp.: 295°C
Initial: 90°C, Hold for 1.0 min., program at 20°C/min. to 205°C
Final: program at 30°C/min. to 275°C
Equil. Time: 1.0 min.
Split Ratio: Splitless
He Flow: 1.0 mL/min. @ 200°C
Septum Purge: 2 mL/min.
Purge Off Time: 1.0 min.
Dwell: 30
Solvent Delay: 3.0 min.
Start Acq.:3.0 min.
Stop Run: 9.00 min.
MSD SIM PROGRAM
DrugIons MonitoredRetention Time
Butalbital 181, 195, 196, 237 4.53 min.
Butabarbital112, 169, 184, 211 4.82 min.
Amobarbital112, 169, 184, 239 5.10 min.
Pentobarbital112, 169, 184, 225 5.27 min.
Secobarbital181, 195, 196, 266 5.52 min.
Phenobarbital117, 175, 232, 260 6.57 min.
Phenobarbital-d5 122, 180, 237, 265 6.56 min.
Diphenylhydantoin118, 194, 203, 280 8.15 min.
HOMEBUTTON - PLEASE CLICK HERE TO RETURN TO HOME PAGE IF NAVIGATION BAR IS NOT AVAILABLE AT LEFT OF SCREEN
GO TO:
GV-65 Ordering Information