![]() |
DETECTABUSE™ GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF BARBITURATES IN URINE BY GC-MS |
Although this method has performed well in our laboratory, it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
|
Revised: March, 2003 Supercedes May, 2001 - Revision addresses stability of TEA in sample elution section |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Pipet 1.0 mL of urine to a 16 x 100
mm disposable borosilicate glass tube. Notes: When adding an internal standard dissolved in an organic
solvent to a urine or blood sample, the solvent volume must not exceed
3% of the buffered sample volume. Higher solvent concentrations may produce
extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup
Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. SAMPLE EXTRACTION - (Please see Notes at end of this section
before proceeding)
1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min.
Note: If liquids do not
elute freely by gravity flow there is probably air trapped within the column
bed or frits. Tapping the column mounting plate onto the vacuum box should
initiate flow. Any columns that have not emptied within 5 or 6 min. may
be induced with a low vacuum from a small vacuum pump.
SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. Note: If liquids do not elute freely by gravity flow, there
is either air trapped within the column bed or frits, or there is aqueous
phase remaining on the column because of a weak vacuum during the column
drying step. Tapping the column mounting plate onto the vacuum box should
initiate flow. If this does not do the job, use a rubber bulb to
gently push a few drops of elution solvent and trapped air into the collection
tube. Allow the remainder of solvent to flow by gravity. 1. Dissolve residue in 100 µL of 0.2M Trimethylphenyl Ammonium Hydroxide
(TMPAH) in Methanol. SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure.
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 280°C |
| Detector Temp.: 295°C |
| Initial: 90°C, Hold for 1.0 min., program at 20°C/min. to 205°C |
| Final: program at 30°C/min. to 275°C |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 3.0 min. |
| Start Acq.:3.0 min. |
| Stop Run: 9.00 min. |
| MSD SIM PROGRAM | ||
| Drug | Ions Monitored | Retention Time |
| Butalbital | 181, 195, 196, 237 | 4.53 min. |
| Butabarbital | 112, 169, 184, 211 | 4.82 min. |
| Amobarbital | 112, 169, 184, 239 | 5.10 min. |
| Pentobarbital | 112, 169, 184, 225 | 5.27 min. |
| Secobarbital | 181, 195, 196, 266 | 5.52 min. |
| Phenobarbital | 117, 175, 232, 260 | 6.57 min. |
| Phenobarbital-d5 | 122, 180, 237, 265 | 6.56 min. |
| Diphenylhydantoin | 118, 194, 203, 280 | 8.15 min. |
![]() GO TO: GV-65 Ordering Information | ||