DETECTABUSE™ "NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF PHENCYCLIDINE (PCP) IN URINE, SERUM OR ORAL FLUID BY GC-MS |
Although this method has performed well in our laboratory, it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: January 2003 (* TEA Elution Solvent Stability) |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Pipet 2.0 mL of sample into a
16 x 100 mm disposable borosilicate glass tube with an inert screw cap top. Notes: When adding an internal standard dissolved in an organic
solvent to a sample, the solvent volume must not exceed 3% of the buffered
sample volume. Higher solvent concentrations may produce extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup
Instructions) >
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. Note: If liquids do not elute freely by gravity flow, there
is probably air trapped within the column bed or frits. Tapping the column
mounting plate onto the vacuum box should initiate flow.
RECONSTITUTION
1. To each dried extract add 100 µL Ethyl Acetate, vortex mix, then
flush with nitrogen or argon. SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure.
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 250°C |
| Detector Temp.: 280°C |
| Initial: 110°C, Hold for 1.0 min., program at 20°C/min. to 210°C |
| Final: program at 30°C/min. to 275°C |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 3.5 min. |
| Start Acq.:3.5 min. |
| Stop Run: 8.25 min. |
| MSD SIM PROGRAM | ||
| Drug | Ions Monitored | Retention Time |
| Phencyclidine | 200, 242, 243 | 5.50 min. |
| Phencyclidine-D5 | 205, 247, 248 | 5.48 min. |
Retention time and ion spectra will vary somewhat from instrument to instrument.![]() GO TO: GV-65 Ordering Information | ||