![]() |
DETECTABUSE™ "NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF LSD AND METABOLITES IN URINE BY GC-MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: January 2003 (* TEA Elution Solvent Stability) |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Add 5.0 mL of urine to a 16 x 100 mm
disposable borosilicate glass tube with an inert screw cap top. Note: When adding an internal standard dissolved in an organic
solvent to a urine or blood sample, the solvent volume must not exceed
5% of the buffered sample volume. Higher solvent concentrations may produce
extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there
is probably air trapped within the column bed or frits. Tapping the column
mounting plate onto the vacuum box should initiate flow.
SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. Note: If a sample does not elute freely by gravity flow, there
is probably air trapped within the column bed or frits. In most cases,
tapping the column will initiate flow. DERIVATIZATION
1. To each dried extract add 50 µL BSTFA and 15 µL TMS-imidazole. PRECAUTION: Derivatized LSD is subject to loss by adsorption
in the injection sleeve and on the capillary column as it ages. When sensitivity
begins to drop, inject a few 5 µL aliquots of Silyl-8 GC column conditioning
solution (Pierce Reagents) onto the capillary column with the oven set
to 150 °C and the purge set to off. If this does not restore sensitivity,
changing the injector sleeve and if necessary, cutting 6-8 inches off
the capillary column should restore it. We suggest that you first run an unextracted standard at the LOQ level
to make sure that the "system" is sensative enough to determine picogram
levels of LSD. SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure.
|
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 285°C |
| Detector Temp.: 300°C |
| Initial: 160°C, program at 20°C/min. to 320°C |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 100 |
| Solvent Delay: 6.5 min. |
| Start Acq.:6.5 min. |
| Stop Run: 9.0 min. |
|
BSTFA/TMSI
|
||
| Drug | Ions Monitored | Retention Time |
| 2-Oxo-3-Hydroxy-LSD | 307, 309, 397 | 8.10 |
| Iso-LSD | 279, 293, 395 | 8.22 |
| LSD | 279, 293, 395 | 8.35 |
| LAMPA | 279, 293, 395 | 8.47 |
Retention time and ion spectra will vary somewhat from instrument to instrument.![]() GO TO: GV-65 Ordering Information | ||