![]() |
DETECTABUSE™ "NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF COTININE IN URINE BY GC-MS |
Although this method has performed well in our laboratory, it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: January 2008 (* Bisufite preparation, washes, drying temp.,GC/MS parameters, Nicotine) |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)
1. Pipet 2.0 mL of urine to a 16
x 100 mm disposable borosilicate glass tube with an inert screw cap top. Notes: When adding an internal standard dissolved in an organic
solvent to a urine or blood sample, the solvent volume must not exceed
3% of the buffered sample volume. Higher solvent concentrations may produce
extraction losses.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup
Instructions) >
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. Note: If liquids do not elute freely by gravity flow, there
is either air trapped within the column bed or frits, or there is aqueous
phase remaining on the column because of a weak vacuum during the column
drying step. Tapping the column mounting plate onto the vacuum box should
initiate flow. If this does not do the job, use a rubber bulb to
gently push a few drops of elution solvent and trapped air into the collection
tube. Allow the remainder of solvent to flow by gravity.
RECONSTITUTION
1. To each dried extract add 100 µL Ethyl Acetate, vortex mix, then
flush with nitrogen or argon. SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure.
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 270°C |
| Detector Temp.: 275°C |
| Level 1: Initial Temp. 80°C, program at 15°C/min. to 125°C |
| Level 2: program at 80°C/min. to 275°C |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 3.5 min. |
| Start Acq.:3.5 min. |
| Stop Run: 9.0 min. |
| MSD SIM PROGRAM | ||
| Drug | Ions Monitored | Retention Time |
| Nicotine | 84, 133, 161 | 3.26 min. |
| Cotinine | 98, 119, 176 | 5.10 min. |
| Cotinine-D3 | 101, 122, 179 | 5.00 min. |
Retention time and ion spectra will vary somewhat from instrument to instrument.![]() GO TO: GV-65 Ordering Information | ||