![]() |
DETECTABUSE™ " NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF METHADONE, EDDP AND METHAQUALONE IN URINE BY GC-MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: January 2003 (* TEA Elution Solvent Stability) |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding) 1. Add 1.0 mL of urine to a 16 x 100 mm disposable borosilicate glass
tube with an inert screw cap top (up to 3 mL may be used). Note: When adding an internal standard dissolved in an organic solvent to a urine or blood sample the solvent volume must not exceed 3% of the buffered sample volume. Higher solvent concentrations may produce extraction losses. HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions) COLUMN CONDITIONING 1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there is probably air trapped within the column bed or frits. Tapping the column mounting plate onto the vacuum box should initiate flow. SAMPLE ELUTION 1. Sample elution is done outside of the vacuum box. Note: If a sample does not elute freely by gravity flow, there is probably air trapped within the column bed or frits or aqueous phase remaining on the column because of weak vacuum during the column drying step. In most cases, tapping the column will initiate flow. If this does not do the job, use a rubber bulb to gently push a few drops of elution solvent and trapped air into the collection tube. Allow the remainder of solvent to flow by gravity. RECONSTITUTION 1. Add 100 µL of n-Butyl Chloride:Ethyl Acetate (1:1) to each dried
extract, vortex mix. SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure.
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 285°C |
| Detector Temp.: 300°C |
| Level 1: 1100C, Hold for 0.5 min., program at 25°C/min. to 210°C Level2: Program from 210°C-250°C at 10°C/min. Level3: Program from 250°C-280°C at 30°C/min. |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 3.5 min. |
| Start Acq.:3.5 min. |
| Stop Run:10.0 min. |
| MSD SIM PROGRAM | ||
| Underivatived | ||
| Drug | Ions Monitored | Retention Time |
| EDDP (Methadone Metabolite) | 262, 276, 277 | 5.70 |
| Methadone | 72, 91, 178 | 6.60 |
| Methaqualone | 233, 235, 250 | 6.72 |
Retention time and ion spectra will vary somewhat from instrument to instrument.![]() GO TO: GV-65 Ordering Information | ||