BCDlogo

DETECTABUSE™ " NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF METHADONE, EDDP AND METHAQUALONE IN URINE BY GC-MS

This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements.

Revised: January 2003 (* TEA Elution Solvent Stability)

SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding)

1. Add 1.0 mL of urine to a 16 x 100 mm disposable borosilicate glass tube with an inert screw cap top (up to 3 mL may be used).
2. Add 150 ng/mL of the appropriate deuterated standard to each sample.
3. Add 1.0 mL of 1 % HCl in H2O per mL of urine. Urine pH should be less than 3.0.
4. If cloudy or precipitated Centrifuge for 3 minutes at 3000 RPM.

Note: When adding an internal standard dissolved in an organic solvent to a urine or blood sample the solvent volume must not exceed 3% of the buffered sample volume. Higher solvent concentrations may produce extraction losses.

HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)

COLUMN CONDITIONING

1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity.
2. Add 1.0 mL of a Sodium Bisulfite solution to each column. Allow to flow by gravity.Prepare by dissolving 25 grams of Sodium Bisulfite in 100 mL of a (1:1) mixture of H2O:0.25M Phosphate Buffer, pH 6.0. Prepare monthly.
3. Proceed to Sample Extraction within 60 min. of column conditioning.

SAMPLE EXTRACTION - (Please see Notes at end of this section before proceeding)

1. Pour samples onto preconditioned column. Allow to flow by gravity. Samples will flow through the column at a rate of 1-2 mL/min.
2. Wash column with 3.0 mL of Deionized Water.
3. Wash column with 2.0 mL of Methanol. Allow the columns to flow by gravity.
4. Wash column with 1.0 mL Ethyl Acetate. Allow the columns to flow by gravity. Proceed to Sample Elution.

Note: If liquids do not elute freely by gravity flow, there is probably air trapped within the column bed or frits. Tapping the column mounting plate onto the vacuum box should initiate flow.

SAMPLE ELUTION

1. Sample elution is done outside of the vacuum box.
2. Place the column mounting plate on the elution rack loaded with an appropriate number of 12 x 75 mm or 15 x 85 mm borosilicate glass test tubes. Make sure that the hole pattern on the plate matches the hole pattern on the rack.
3. Add 2.0 mL of Ethyl Acetate w/4% Triethylamine (TEA)* to each column and allow solvent to flow through the columns by gravity into the test tubes.
4. Dry under N2 or argon at less than 50°C.

* Elution solvent with 4% TEA, (4 mL TEA is added to 96 mL of Ethyl Acetate) is stable for approximately one week stored in a glass bottle with a Teflon or polypropylene lined cap. Close bottle tightly when not in use. A white residue begins to appear in the dried down eluate when the TEA begins to deteriorate. Artifacts from this process may interfere with "fast" GC/MS methods.

Note: If a sample does not elute freely by gravity flow, there is probably air trapped within the column bed or frits or aqueous phase remaining on the column because of weak vacuum during the column drying step. In most cases, tapping the column will initiate flow. If this does not do the job, use a rubber bulb to gently push a few drops of elution solvent and trapped air into the collection tube. Allow the remainder of solvent to flow by gravity.

RECONSTITUTION

1. Add 100 µL of n-Butyl Chloride:Ethyl Acetate (1:1) to each dried extract, vortex mix.
2. Flush with Nitrogen or Argon and cap the tube or transfer contents into 100 µL reaction vials and seal.
3. Inject 1-2 µL.

SUPPLEMENT - When using an automated robotic system all liquids may be allowed to flow unassisted through the column or may be pulled through the column with vacuum or pushed through with positive pressure.
Assisted flow parameters may be set as follows:
Column Conditioning - Pass through column in approximately 20 seconds (± 20%).
Sample, Sample Washes, and Elution Solvent - Pass through column in approximately 60 seconds (± 20%).


GC/MS ANALYSIS
GC/MS: Hewlett-Packard equipped with Mass Selective Detector
GC Column: H.P. Ultra 2 Capillary Column (or equivalent),
15 m x 0.25 mm, 0.25 µm film
Acquisition Mode: SIM

Temperature Program:
Injector Temp.: 285°C
Detector Temp.: 300°C
Level 1: 1100C, Hold for 0.5 min., program at 25°C/min. to 210°C
Level2: Program from 210°C-250°C at 10°C/min.
Level3: Program from 250°C-280°C at 30°C/min.
Equil. Time: 1.0 min.
Split Ratio: Splitless
He Flow: 1.0 mL/min. @ 200°C
Septum Purge: 2 mL/min.
Purge Off Time: 1.0 min.
Dwell: 30
Solvent Delay: 3.5 min.
Start Acq.:3.5 min.
Stop Run:10.0 min.
MSD SIM PROGRAM
Underivatived
DrugIons MonitoredRetention Time
EDDP (Methadone Metabolite) 262, 276, 277 5.70
Methadone 72, 91, 178 6.60
Methaqualone 233, 235, 250 6.72
Retention time and ion spectra will vary somewhat from instrument to instrument.
HOMEBUTTON - PLEASE CLICK HERE TO RETURN TO HOME PAGE IF NAVIGATION BAR IS NOT AVAILABLE AT LEFT OF SCREEN
GO TO:
GV-65 Ordering Information