![]() |
DETECTABUSE™ "NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF AMPHETAMINE, METHAMPHETAMINE, 3,4 METHYLENEDIOXYAMPHETAMINE (MDA), 3,4 METHYLENEDIOXYMETHAMPHETAMINE (MDMA), 3,4 METHYLENEDIOXYETHYLAMPHETAMINE (MDEA) IN ORAL FLUID, SERUM OR URINE BY GC/MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory it must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: April 2004 (HFBA Incubation change) |
|
NOTE: This revision eliminates variability caused by incubating the HFBA derivatives at room temperature. Room temperature incubation works when drydown is done at 60°C because sufficient energy is provided for derivatization to take place during the drydown process. Drydown is usually done at much lower temperatures to prevent evaporative losses and this may introduce an unacceptable variability in derivatization results. |
| SAMPLE PREPARATION WITHOUT
PERIODATE OXIDATION - (Please see Notes and Supplemental Information
before proceeding)
1. Add 1.0 mL of sample to a 16 x 100 mm disposable borosilicate glass
tube with an inert screw cap top. Note: When adding an internal standard dissolved in an organic solvent to a sample, the solvent volume must not exceed 5% of the buffered sample volume. Higher solvent concentrations may produce extraction losses.
WITH PERIODATE OXIDATION (OPTIONAL) 1. Add 2.0 mL of 0.25 M Phosphate
Buffer, pH 9.1 to 1 mL of sample spiked with internal standard as above. Preparation of Sodium Metaperiodate:
Add 10 grams of Sodium Metaperiodate to 100 mL of deionized H2O, mix well
until solution is clear. 3. Incubate at room temperature for 20 min. HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there is
probably air trapped within the column bed or frits. Tapping the column
mounting plate onto the vacuum box should initiate flow.
SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. DERIVATIZATION - Following are some of the more commonly used
derivatizing schemes. Others such as PFPA are available upon request.
Using HFBA Using the 4-CB Derivative: Note: Polar solvents commonly used for derivatization
such as Acetonitrile or Ethyl Acetate will pick up moisture over time.
Because moisture will inhibit or prevent derivatization it is important
to keep a supply of solvents used for this purpose stored as protected
from moisture as possible. Storing these solvents in a desiccator or flushing
with Nitrogen or Argon after use is recommended.
SUPPLEMENT - When using an automated robotic system
All liquids may be allowed to flow unassisted through the column or
may be pulled through the column with vacuum or pushed through with positive
pressure.
|
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.:175°C |
| Detector Temp.: 250°C |
| Using HFBA Initial: 75°C, Hold for 0.5 min., Increase at 20°C/min. to 200°C Final: Increase from 200°C-275°C at 40°C/min. Using 4-CB Initial: 120°C , Increase at 20°C/min. to 225°C Final: Increase to 275 °C at 20°C/min. |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 3.0 min. |
| Start Acq.:3.0 min. |
| Stop Run: 9.0 min. |
.![]() GO TO: GV-65 Ordering Information |
| MSD SIM PROGRAM | ||||
| HFBA | 4-CB | |||
| Drug | Ions Monitored | Retention Time | Ions Monitored | Retention Time |
| Amphetamine-D5 | 123, 244 | 3.50 min. | 298, 270 | 3.90 min. |
| Amphetamine | 91, 118, 240 | 3.55 min. | 294, 266, 248 | 3.93min. |
| Methamphetamine-D5 | 213, 258 | 4.19 min. | 312, 284 | 4.36 min. |
| Methamphetamine | 118, 210, 254 | 4.21 min. | 308, 280, 262 | 4.39 min. |
| MDA-D5 | 167, 270 | 5.50 min. | 167, 270 | 5.56 min. |
| MDA | 162, 248, 266 | 5.53 min. | 162, 248, 266 | 5.59 min. |
| MDMA-D5 | 213, 258 | 6.12 min. | 167, 284 | 5.95 min. |
| MDMA | 162, 210, 254 | 6.15 min. | 162, 280, 308 | 5.98 min. |
| MDEA-D5 | 274, 409 | 6.71 min. | ||
| MDEA | 162, 240, 268, 403 | 6.77 min. | ||
| Retention time and ion spectra will vary somewhat from instrument to instrument | ||||