|
DETECTABUSE™ "NO VACUUM" GRAVITY SERIES GV-65 METHOD FOR THE ANALYSIS OF 6-MONOACETYLMORPHINE IN URINE, SERUM OR ORAL FLUID BY GC-MS |
This method is a preliminary procedure for investigational use only. Although it has performed well in our laboratory, the method must be validated by your laboratory before it is used to report patient values. We would appreciate your comments on its performance and welcome your suggestions for improvements or enhancements. |
Revised: January 2003 (* TEA Elution Solvent Stability) |
SAMPLE PREPARATION - (Please see Notes and Supplemental Information before proceeding) 1. Add 3.0 mL of sample to a
16 x 100 mm disposable borosilicate glass tube with an inert screw cap
top. Note: a. When adding an internal standard dissolved in an
organic solvent to a sample, the solvent volume must not exceed 3% of
the buffered sample volume. Higher solvent concentrations may produce
extraction losses.
Note: b.
The Sodium Bisulphite reduction step, when combined with the unique derivatization
step, eliminates the interference from Codones and Morphones without affecting
morphine, codeine or 6-MAM . If this interference is not a concern, this
reduction step may be eliminated. The column still must be pretreated
with Sodium Bisulphite to activate the cation exchange moiety on the resin.
HARDWARE SETUP - (Please refer to the Detectabuse Hardware Setup
Instructions)
COLUMN CONDITIONING
1. Wash column with 1.0 mL of Methanol. Allow to flow by gravity. 1. Pour samples onto preconditioned column. Allow to flow by gravity.
Samples will flow through the column at a rate of 1-2 mL/min. Note: If liquids do not elute freely by gravity flow, there
is probably air trapped within the column bed or frits. Tapping the column
mounting plate onto the vacuum box should initiate flow.
SAMPLE ELUTION
1. Sample elution is done outside of the vacuum box. Note: If a sample does not elute freely by gravity flow,
there is probably air trapped within the column bed or frits. In most
cases, tapping the column will initiate flow. If this does not do the
job, use a rubber bulb to gently push a few drops of elution solvent and
trapped air into the collection tube. Allow the remainder of solvent to
flow by gravity. DERIVATIZATION - Using TMSI/MBTFA
1. To each dried extract add 70 µL Acetonitrile and 15 µL Trimethylsilylimidizole (TMSI).
SUPPLEMENT - When using an automated robotic system all liquids
may be allowed to flow unassisted through the column or may be pulled
through the column with vacuum or pushed through with positive pressure. |
GC/MS ANALYSIS |
| GC/MS: Hewlett-Packard equipped with Mass Selective Detector |
| GC Column: H.P. Ultra 2 Capillary Column (or equivalent), 15 m x 0.25 mm, 0.25 µm film |
| Acquisition Mode: SIM |
| Temperature Program: |
| Injector Temp.: 285°C |
| Detector Temp.: 295°C |
| Initial: 110°C, program at 25°C/min. to 300°C |
| Equil. Time: 1.0 min. |
| Split Ratio: Splitless |
| He Flow: 1.0 mL/min. @ 200°C |
| Septum Purge: 2 mL/min. |
| Purge Off Time: 1.0 min. |
| Dwell: 30 |
| Solvent Delay: 5.0 min. |
| Start Acq.: 5.0 min. |
| Stop Run: 9.0 min. |
| MSD SIM PROGRAM | ||
| TMSI/MBTFA | ||
| Drug | Ions Monitored | Retention Time |
| 6-Monoacetylmorphine | 340, 399, 400 | 7.33 |
| 6-Monoacetylmorphine-d3 | 290 ,343, 402 | 7.32 |
| 6-Monoacetylmorphine-d6 | 290, 343, 405 | 7.31 |
Retention time and ion spectra will vary somewhat from instrument to instrument.![]() GO TO: GV-65 Ordering Information | ||